Multipara Monitor - Comen Star 8000 is conventional equipment that finds its application in Arrhythmia analysis and S-T segment analysis ensuring high level of accuracy. This ensures the ECG waveforms of 7-leads display on the same screen. Suitable for adult, pediatric and neonatal patient this equipment is networkable with Contec central monitoring system.
Features:
- 12.1 inch high resolution color TFT display
- Unique NIBP button and NIBP dust shield Design
- Support Nurse Call connection
- Fully meeting basic monitoring requirement
Patient Range:
- Adult, Pediatric, and Neonate patients
Performance Specifications:
Display:
- 12.1inch color TFT (diagonal)
- Resolution: 800x600
Trace:
- 5 waveforms
- Sweep Speed: 12.5mm/s, 25mm/s, 50mm/s
Indicator:
- Power indicator light
- Battery indicator light
- Alarm indicator light
- QRS beep and alarm sound
Inter face:
- Parameter cable inter face
- AC Power input socke
- Network inter face
Battery:
- lead-Acid /Rechargeable
- 5 hours for continuous working
Trend time:
- 1~96 hours
Alarm:
- User-adjustable High and Low limits
- alarm events recallable
- audible and visual alarm
Networking:
- Connected to central monitoring system
Recorder:
- Built-in, thermal array,
- 2 traces, paper width:50mm
- Record mode: manual, on alarm,
- time -defined
Print speed:
- 12.5 mm/s, 25mm/s, 50mm/s
Multipara Monitor COMEN STAR 8000
SKU : YE-ICU-COMEN STAR 8000
900,00$Prix
ECG
- Lead selection: 5-Lead and 3-lead selectable
- 5 leads RA; LA; RL; LL; V or R; L; N; F; C,
- 3 leads: RA; LA; LL or R; L; F
- ECG waveform: 2 channels
- Gain selection : 1/4 1/2 1 2 Auto
- Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
- Heart Rate Range: 15 ~ 350bpm
- Accuracy: 1% or 2bpm whichever is greater
- Resolution: 1 bpm
- Waveform: Standard 5 ECG waveforms
- Pace detection: Yes
- HR alarming time: 12s
- Input impedance: > 5 M
Respiration:
- Method: Thoracic impedance
- RR measurement range: 0-100rpm
- Resolution: 1 rpm
- Accuracy: 1rpm
SpO2
- Measurement range: 0-100%
- Resolution: 1%
- Accuracy: 1% (90-100%) 2% (70-89%)
- Pulse Rate range 0~250 bpm
- Pulse Rate Accuracy: 1bpm
NIBP
- Measurement method: Automatic oscillometric
- Measurement types: Systolic, Diastolic, Mean
- Work mode: Manual / Automatic
- Auto measurement time: Adjustable
- Measurement unit: mmHg / KPa selectable
Range of systolic pressure:
- Adult mode: 40~270mmHg
- Pediatric mode: 40~200:mmHg
- Neonate mode: 25~135 mmHg
Range of diastolic pressure:
- Adult mode: 30~200mmHg
- Pediatric mode:20~150 mmHg
- Neonate mode: 10~110mmHg
Range of mean pressure:
- Adult mode: 30~220mmHg
- Pediatric mode 30~160mmHg
- Neonate mode: 20~170 mmHg
- Accuracy: ±5mmHg
- NIBP pressure range 0~300 mmHg
- PR from NIBP: Measurement 40-240bpm
- Over-pressure protection: Double safety protection
Temperature
- Measurement range: 20-45°C
- Resolution: 0.1°C
- Accuracy: ±0.1°C
- Measurement unit: °C/F selectable
Sample Gas Flowrate: 50ml/mi